Monitoring invasive pressure, Intended use, Indications – Physio-Control LIFEPAK 15 User Manual

Page 98: Contraindications, Ip monitoring warnings

Advertising
background image

4-54

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring Invasive Pressure

Monitoring Invasive Pressure

Intended Use

The LIFEPAK 15 invasive pressure (IP) monitor is intended for measuring arterial, venous,

intracranial, and other physiological pressures using an invasive catheter system with a compatible

transducer.

The IP monitor is a tool to be used in addition to patient assessment. Care should be taken to

assess the patient at all times; do not rely solely on the IP monitor.

Indications

Invasive pressure monitoring is indicated for use in patients who require continuous monitoring

of physiological pressures in order to rapidly assess changes in the patient’s condition or

response to therapy. It may also be used to aid in medical diagnosis.

Contraindications

None known.

IP Monitoring Warnings

WARNINGS

INACCURATE READINGS HAZARDS

POSSIBLE INACCURATE PRESSURE READINGS, AIR EMBOLISM, BLOOD LOSS, OR LOSS OF
STERILITY

Before use, carefully read these operating instructions, and the transducer and

infusion set instructions for use and precautionary information.

INACCURATE PRESSURE READINGS

Pressure readings should correlate with the patient’s clinical presentation. If readings

do not correlate, verify that the zeroing stopcock is positioned at the patient’s zero

reference, rezero the transducer, and/or check the transducer with a known or

calibrated pressure. Manually check cuff blood pressure.

INACCURATE PRESSURE READINGS

Changing the patient’s position changes the zero reference level. Relevel the

transducer’s zeroing stopcock any time the patient’s position is changed.

Advertising