Acquiring a 12-lead ecg – Physio-Control LIFEPAK 15 User Manual

Page 68

Advertising
background image

4-24

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Acquiring a 12-Lead ECG

For general troubleshooting tips, see Table 10-2 on page 10-18.

Monitor does not
complete 12-lead
ECG operation
sequence

Operator pressed another
function button (such as
PRINT

) before 12-lead

ECG sequence completed

• Press 12-LEAD to acquire another 12-lead

ECG. Allow enough time for sequence to
complete.

Noisy signal and
message displayed:
EXCESSIVE NOISE–
12-LEAD CANCELLED

Signal noise for more than
30 seconds

• Press 12-LEAD to acquire another 12-lead

ECG.

Baseline wander (low
frequency/high
amplitude artifact)

Inadequate skin
preparation

• Prepare skin as described on page 4-8 and

apply new electrodes.

Poor electrode-skin
contact

• Check electrodes for proper adhesion.

Fine baseline artifact
(high frequency/low
amplitude)

Inadequate skin
preparation

• Prepare skin as described on page 4-9 and

apply new electrodes.

Isometric muscle tension
in arms/legs

• Confirm that limbs are resting on a

supportive surface.

• Check electrodes for proper adhesion.

Table 4-3 Troubleshooting Tips for the 12-Lead ECG (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION