10-6

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

General Maintenance and Testing

Note:

It is important to understand defibrillator operation. For suggested procedures to help keep

personnel acquainted with normal defibrillator operation, see the function checks that are provided

in this chapter. The function checks used may vary according to your local protocols. To test the

defibrillator by performing the function checks, you need a simulator. To troubleshoot device

performance, see Table 10-2 on page 10-18.

Standard Paddles User Test

Perform the Standard Paddles User Test as a part of completing the daily Operator’s Checklist that

is provided in the back of this manual.

Function Checks

The following function checks are provided to help personnel keep acquainted with normal

operating procedures and to troubleshoot LIFEPAK 15 monitor/defibrillator performance.

Note:

If your organization downloads device electronic patient records for post-event review,

consider entering “TEST” as the patient’s name to distinguish simulator function tests from actual

patient uses.

Patient ECG Cable Check

Equipment Needed:

• LIFEPAK 15 monitor/defibrillator

• Fully charged batteries or power adapter connected to a reliable power source

• Patient ECG cable (3-lead, 12-lead, or 5-wire)

• 3-lead or 12-lead simulator

To check the patient ECG cable:

1. Press

ON

.

2. Connect the ECG cable to the defibrillator.

3. Connect all cable leads to the simulator.

4. Turn on the simulator and select a rhythm.

5. Confirm that Lead

II

is selected.

6. After a few seconds, confirm that the screen displays a rhythm and that no

LEADS OFF

or

SERVICE

message appears.

7. For 12-lead cable, press

12-LEAD

and wait for printout. Confirm that a rhythm prints for each

lead.