Appendix – Physio-Control LIFEPAK 15 User Manual

Page 241

Advertising
background image

APPENDIX

A

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

A-7

TEMPERATURE

Sensors

Measurement Specialties 4400 series esophageal/rectal and Foley
catheter temperature probes, and 4499HD skin temperature probe

Displayed Range

24.8° to 45.2°C (76.6° to 113.4°F)

Resolution

0.1°C

Accuracy

±0.2°C

Labels

Temp, T-esoph, T-naso, T-bladder, T-rectal, T-skin

Update Rate

Every 10 seconds, minimum

Adapter Cable

Only use Physio-Control part number 3303935

Cable Length

1.5 or 3 m (5 or 10 ft)

TREND

Time Scale

Auto, 30 minutes, 1, 2, 4, or 8 hours

Duration

Up to 8 hours

ST

After initial 12-lead ECG analysis, automatically selects and trends ECG
lead with the greatest ST displacement

Display

Choice of HR, PR (SpO

2)

, PR (NIBP), SpO

2

(%), SpCO(%), SpMet(%),

CO

2

(EtCO

2

/FiCO

2)

, RR (CO

2

), NIBP, IP1, IP2, TEMP, ST

ALARMS

Quick Set

Activates alarms for all active vital signs

VF/VT Alarm

Activates continuous CPSS monitoring in Manual mode

Apnea Alarm

Occurs when 30 seconds has elapsed since last detected respiration

Heart Rate Alarm Limit
Range

Upper, 100–250 bpm; lower, 30–150 bpm

INTERPRETIVE
ALGORITHM

12-Lead Interpretive Algorithm: University of Glasgow 12-Lead ECG
Analysis Program, includes AMI and STEMI statements

PRINTER

Prints continuous strip of the displayed patient information and reports

Paper Size

100 mm (3.9 in)

Print Speed

25 mm/sec or 12.5 mm/sec
Optional 50 mm/sec time base for 12-lead ECG reports

Delay

8 seconds

Autoprint

Waveform events print automatically

Frequency Response

Diagnostic—0.05 to 150 Hz or 0.05 to 40 Hz
Monitor—0.67 to 40 Hz or 1 to 30 Hz

Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)

CHARACTERISTIC

DESCRIPTION