How capnography works, Monitoring etco2 – Physio-Control LIFEPAK 15 User Manual

Page 90

Advertising
background image

4-46

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring ETCO2

How Capnography Works

An EtCO

2

sensor continuously monitors carbon dioxide (CO

2

) that is inspired and exhaled by the

patient. The sensor employs Microstream non-dispersive infrared (IR) spectroscopy to measure the

concentration of CO

2

molecules that absorb infrared light.

The CO

2

FilterLine system delivers a sample of the exhaled gases directly from the patient into the

LIFEPAK 15 monitor for CO

2

measurement. The low sampling flow rate (50 ml/min) reduces liquid

and secretion accumulation and prevents obstruction, which maintains the shape of the CO

2

waveform.

The CO

2

sensor captures a micro sample (15 microliters). This extremely small volume allows for

fast rise time and accurate CO

2

readings, even at high respiration rates.

The Microbeam IR source illuminates the sample cell and the reference cell. This proprietary IR

light source generates only the specific wavelengths characteristic of the CO

2

absorption spectrum.

Therefore, no compensations are required when concentrations of O

2

, anesthetic agent, or water

vapor are present in the exhaled breath.

INACCURATE READINGS HAZARDS

POSSIBLE INACCURATE PATIENT ASSESSMENT

The EtCO

2

monitor is intended only as an adjunct in patient assessment and is not to

be used as a diagnostic apnea monitor. An apnea message appears if a valid breath

has not been detected for 30 seconds and indicates the time elapsed since the last

valid breath. It must be used in conjunction with clinical signs and symptoms.

POSSIBLE INACCURATE CO2 READINGS

Using other manufacturers’ CO

2

accessories may cause the device to perform

improperly and invalidate the safety agency certifications. Use only the accessories

that are specified in these operating instructions.

HEALTH HAZARDS

POSSIBLE STRANGULATION

Carefully route the patient tubing (FilterLine) to reduce the possibility of patient

entanglement or strangulation.

INFECTION HAZARD

Do not reuse, sterilize, or clean Microstream

®

CO

2

accessories as they are designed for

single-patient one-time use.

WARNINGS (CONTINUED)

Advertising