Specifications and performance characteristics – Physio-Control LIFEPAK 15 User Manual

Page 236

Advertising
background image

A-2

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Specifications and Performance Characteristics

Analog ECG Output

Output: 1 volt/mV
Frequency Response: 0.67 to 32 Hz (except 2.5 to 25 Hz for Paddles
ECG and 1.3 to 23 Hz for 1–30 Hz Monitor Frequency Response)

Notch Filter

50 or 60 Hz

POWER

Batteries

Rechargeable Lithium-ion battery

Dual battery capability with automatic switching

Low battery indication and message: Low battery fuel gauge indication
and low battery message in status area for each battery

Replace battery indication and message: Replace battery fuel gauge
indication, audio tones, and replace battery message in the status area
for each battery. When replace battery is indicated, device auto-switches
to second battery. When both batteries reach replace battery condition, a
voice prompt instructs user to replace battery.

Input voltage range is between +8.8 and +12.6 Vdc

Battery Capacity

For two, new fully-charged batteries, 20°C (68°F):

Capacity to shutdown is:

Operating Mode

Monitoring (minutes)

Pacing (minutes)

Defibrillation

(360J discharges)

Typical

360

340

420

Minimum

340

320

400

Capacity after low battery is:

Typical

21

20

30

Minimum

12

10

6

AC Power Adapter

AC-DC power adapter

Input power range is 100-240 Vac, 50/60 Hz, 1.4-0.6 A

Output voltage is 12 Vdc

Meets UL 60601-1 300 microampere Earth leakage limit when installed
on a center-tapped, 240 Vac, single phase circuit.

DC Power Adapter

DC-DC power adapter

Input power range is 11 to 17.6 Vdc, 15 A

Output voltage is 12 Vdc

Device Behavior

when using

Power Adapter

Auxiliary power indicator on defibrillator illuminated when connected to
auxiliary power. Battery charging indicator illuminated when batteries are
fully charged and flashing if either battery is being charged.

Battery status indicators on display show battery charge level, but well
number is not highlighted because battery is not in use. Low battery and
replace battery prompts and messages do not occur.

Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)

CHARACTERISTIC

DESCRIPTION