General troubleshooting tips – Physio-Control LIFEPAK 15 User Manual

Page 226

Advertising
background image

10-20

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

General Troubleshooting Tips

Fuel gauge on battery does
not illuminate

Extremely depleted battery

• Charge battery in Station-Mobile or

REDI-CHARGE battery charger.

Faulty battery

• Replace battery.

Device turns off
unexpectedly

High power draw

• Press ON immediately to turn device

back on.

Low battery power

• Replace battery immediately.
• Press ON to turn device back on.

RF equipment too close to
defibrillator

• Separate RF equipment from

defibrillator. See "Separation
Distances" on page D-4.

• Press ON to turn device back on.

Cellular equipment too close
to installed battery

• Move cellular equipment away from

installed battery.

• Press ON to turn device back on.
• If device does not turn on, replace

battery.

LIFENET Gateway (modem)
too close to installed battery

• Store modem in side pouch of

defibrillator. Do not store modem in
back pouch.

• Press ON to turn device back on.
• If device does not turn on, replace

battery.

Device won’t turn off

ON

not pressed long enough

to turn off device

• Press and hold ON for at least two

seconds.

Monitor/defibrillator
operates, but screen is
blank

Operating temperature is too
low or too high

• Operate defibrillator within specified

ambient temperature range.

Screen not operating
properly

• Print ECG strip to assess rhythm and

other active vital signs.

• Press ANALYZE and use AED mode, if

necessary.

• Contact qualified service technician.

Monitor/defibrillator
operates, but screen not
readable

Screen in direct sunlight

• Change screen from color to black

and white.

• Reposition or shield device.
• Print ECG strip to assess rhythm and

other active vital signs.

• Press ANALYZE and use AED mode, if

necessary.

Table 10-2 General Troubleshooting Tips (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION