Table a-3 list, A-14, Specifications and performance characteristics – Physio-Control LIFEPAK 15 User Manual

Page 248

Advertising
background image

A-14

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Specifications and Performance Characteristics

Table A-3 Alarm Limits

VITAL SIGN
(VS)

PATIENT
VS VALUE

WIDE LIMITS*

NARROW LIMITS*

LIMITS RANGE

DEFAULT LIMITS**

LOW

HIGH

LOW

HIGH

LOW

HIGH

LOW

HIGH

Heart Rate

(HR)

Pulse Rate
(PR)

<60

-20

+35

-10

+25

30–150 100–250 50

150

60–79

-25

+40

-20

+30

80–104

-30

+40

-30

+30

(bpm)

>105

-35

+45

-25

+25

SpO

2

(%)

>90

-5

+3

-5

+3

50

90–100

85

100

<90

-5

+3

-5

+3

Systolic BP
(mmHg)

<90

-20

+35

-10

+25

30

245

50

200

90–114

-20

+35

-10

+25

115–140

-25

+35

-10

+20

>140

-25

+35

-10

+20

Diastolic BP
(mmHg)

<65

-15

+25

-10

+25

12

210

20

150

65-90

-15

+15

-15

+10

>90

-15

+15

-15

+10

EtCO

2

(mmHg/%)

>40/5.3

-10/-1.3 +15/+2.0 -10/-1.3 +15/+2.0 5/0.7

70/9.2

15/2.0

50/6.6

40/5.3

-10/-1.3 +15/+2.0 -10/-1.3 +15/+2.0

Inspired CO

2

(mmHg/%)

n/a

+5/+0.7

n/a

+3/+0.4

n/a

0/0–10/
1.3

8/1.1

Respiration

Rate (RR)

<15

-8

+8

-4

+4

5-15

10–60

5

30

15

-15

+15

-8

+8

Systolic PA
(mmHg)

<15

-6

+12

-4

+6

10

100

10

40

15–35

-8

+16

-6

+8

>35

-12

+16

-8

+10

Diastolic PA
(mmHg)

<5

-4

+12

-4

+8

0

50

0

18

5–13

-4

+16

-6

+6

>13

-6

+16

-6

+6

CVP (mmHg)

9

-10

+10

-5

+5

0

25

0

15

ICP, LAP
(mmHg)

<15

-6

+6

-4

+4

0

40

0

18

15

-6

+8

-4

+6

Temperature
(°C)



-3

+3

-1

+1

31

41

35

39

*Numbers are ± from patient’s VS value when the alarms are set.

**Default limits are established when alarms are set up to be ON.