Screen messages – Physio-Control LIFEPAK 15 User Manual

Page 258

Advertising
background image

B-4

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Screen Messages

MAINTENANCE DUE

Reminder message that appears at the interval that is set in
Service mode. Message continues to appear until reset or
turned off.

MANUAL MODE DISABLED

Access to Manual mode from AED mode has been restricted.

MOTION DETECTED!/STOP
MOTION!

Motion was detected during ECG analysis.

NIBP AIR LEAK

NIBP cuff applied too loosely or there is a leak in cuff/monitor
pneumatic system.

NIBP CHECK CUFF

NIBP cuff is not connected to patient or device.

NIBP FAILED

NIBP monitor cannot establish zero-pressure reference.

NIBP FLOW ERROR

NIBP pneumatic system is not maintaining stable cuff pressure.

NIBP INITIALIZING

NIBP requested while NIBP module is still initializing.

NIBP MOTION

Patient extremity moved too much for the NIBP monitor to
accurately complete the measurement.

NIBP OVERPRESSURE

NIBP cuff pressure exceeded 290 mmHg.

NIBP TIME OUT

NIBP monitor did not complete a measurement in 120 seconds.

NIBP WEAK PULSE

The monitor did not detect any pulses.

NO SHOCK ADVISED

The defibrillator did not detect a shockable rhythm.

NO SITES DEFINED

Device is attempting to transmit using Bluetooth connection,
but no associated destinations have been defined.

NOISY DATA! PRESS 12-LEAD TO
ACCEPT

Monitor detects excessive signal interference while acquiring
data. Press 12-LEAD to override the message and acquire 12-
lead ECG with noise.

NON-DEMAND

Pacemaker is in Nondemand (asynchronous) mode.

OBTAINING DEVICE NAMES

Device is obtaining names of available Bluetooth-enabled
devices.

PACER FAULT

Internal error detected during pacing.

PACING IN PROGRESS

The requested action is not available because the device is
currently performing pacing.

PACING STOPPED

Pacing has stopped—for example, due to disconnection of
therapy electrodes.

PASSCODE INCORRECT -
TRY AGAIN

Incorrect passcode entered.

PAUSED

The pacing PAUSE button is pressed and held. Current pulses
are applied at reduced frequency while the MA and PPM
settings are maintained.

Table B-1 Summary of Screen Messages (Continued)

MESSAGE

DESCRIPTION