3

Operation

PC Control / Touch Control

91

Enter either the Validation interval or the date for the Next

validation. If the time interval or date is edited then the other

parameter will be updated automatically.

Select an Action to be carried out when the time interval has

expired.

The validation interval will always be monitored at the START of a

determination.
You can enter a short note under System validation/Note (SOP),

e. g. a summary of the SOP (standard operating procedure) according

to which the system validation was carried out.

3.12.5 Service interval

You can monitor the time interval after which the Titrando system must

be serviced again.

Choose [Service interval] in the dialog GLP manager/Monitor-

ing. Enter the date upon which the last service was carried out and

activate the Monitoring checkbox.

Enter either the Service interval or date for Next service. If the

time interval or date is edited then the other parameter will be

updated automatically.

When the service interval has expired then the corresponding message

will appear at the START of a determination; this will also be

documented in the determination data (see Section 3.18.2).

3.12.6 Backup interval

You can monitor the time interval after which a backup should be

carried out.

Choose [Backup interval] ] in the dialog GLP manager/Monitor-

ing. Activate the Monitoring checkbox.

Enter either the Backup interval or the date for the next backup.

If the time interval or date is edited then the other parameter will be

updated automatically.

When the backup interval has expired a corresponding message will

appear when the system is started.

Advertising