Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 215

Advertising
background image

4

Parameters

PC Control / Touch Control

203

V

E

Inflection

point

pH

V/mL

Fig. 4.7: Tubbs method for determining the equivalence point

The figures shows that the evaluation also requires measuring points

after the equivalence point too.
For the recognition of the EPs found the set EP criterion is compared

with the ERC (Equivalence point Recognition Criterion) found. The ERC

is the first derivative of the titration curve combined with a mathematical

function which is more sensitive for flat jumps than for steeper ones.

EPs whose ERC is smaller than the set EP criterion will not be recog-

nized. The ERC for each found and recognized EP will be shown under

Results/Curve/EP Details

. If you subsequently adapt the EP criterion

so that more or fewer EPs are recognized, you can trigger the re-

evaluation

under Results with [Recalculate].

For dynamic equivalence point titrations you can also show the ERC in

addition to the titration curve. If the difference between the ERCs of the

endpoints is large enough you will also be able to read off a sensible

value for the EP criterion from the curve.