Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 271

Advertising
background image

4

Parameters

PC Control / Touch Control

259

Unit Conc. (d): Select the concentration unit of the standard. The

concentration unit for concentration measurements always depends

on the unit entered for the concentration of the standard.

Conc. standard 1 to Conc. standard 5 (d): Enter the concentra-

tion for 1 to 5 concentration standards. off means that the standard

and all those following will not be used. During the sequence a

message will appear the corresponding number of times to indicate

the next standard to be used.

Under Stirrer control you can edit the parameters for controlling the

stirrer during the calibration. Take care that you have selected a stirrer

under Edit command/Stirrer.
Stir solution during measurement (d): Automatically switches

the stirrer on/off at the start and the end of the measurement.

Stir before meas. (d): If stirring is switched off during the meas-

urement, the solution can be stirred before the measurement. Enter

the corresponding time here.

Pause before meas. (d): If stirring is carried out before the meas-

urement, a pause interval can be entered here. No stirring or meas-

uring occurs during this time.

The parameters for Control device, Sensor and Stirrer are de-

scribed in Section 4.1.5.

Note!

Make sure that the sensor to be calibrated has been selected under

Edit command/Sensor. This is the only way to ensure that the cali-

bration data for this sensor is stored in the correct place under Sys-

tem/Sensors.

The choice of the Direct parameters is described in Section 4.1.6.

2-Point calibration with automatic buffer recognition

Immerse electrode

in first buffer

With automatic buffer recognition the order of the

buffers does not matter.

Start the

calibration com-

mand

Measure or enter

the calibration tem-

perature

If a temperature sensor is connected then the

temperature will be measured. Otherwise a query

will appear and the temperature must be entered

manually.

(Stir and/or

pause before

measurement)

The solution is stirred before the measurement.

Pause will be allowed to elapse before the meas-

urement.