3.18 Results and more determination data

130

PC Control / Touch Control

3.18.1 More determination data

Under Results/More determination data a list containing all the

method commands which produce raw data (titrations, measurements,

calibrations and monitored dosing) and evaluations (EVAL commands)

is shown.

Open this list with [More determination data].

Under View data the details (raw data and variables) for the selected

command are shown.

Select a command and press [View data].

In the first line the command and the command identification are shown

(in the above example 1M, i.e. the first titration/measurement

command). The command identification is used for the unambiguous

assignment of the variables to a mode. If only one titration or

measurement is carried out in a determination then the command

identification does not need to be given.
Below this is a list of raw data and variables. The abbreviation of the

variable is shown for variables which can be used in calculations. The

individual variables are described in Section 4.5.7. In addition to the

variables, the Stop criterion used to stop the titration, measurement,

etc. (manual stop, stop volume reached, error, etc.) and the type of

temperature measurement (Pt1000, NTC or manual) are shown.