Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 208

Advertising
background image

4.1 Titrations

196

PC Control / Touch Control

Endpoint titrations while holding the measured value con-

stant. pH STAT is the most frequently used titration. A reagent is

used to dose to a preset measured value (control point). This con-

trol point is kept constant by titrating the reaction product continu-

ously with the reagent. The STAT titration keeps the control point

constant until the preset stop condition has been reached. The

STAT titration is used, for example, in enzymatic analysis. The dos-

ing rate resulting from holding the control point constant provides

information about enzyme activity.

Control range

Control

point

V/mL

t/s

U/mV

V/mL

Fig. 4.5: Reagent addition for STAT

pH and U are the only measuring modes that are available for the

STAT titration.