3.8 Titrants

48

PC Control / Touch Control

Select the Dosing port via which the first cylinder contents are to

be ejected during preparation and emptying.

Enter the Dosing rates for all the ports to be used during the

preparation and emptying functions for drawing in and ejecting the

reagent. Lower rates should be entered for substances with higher

viscosities.

The maximum dosing rate and maximum filling rate for the dosing

unit depend on the cylinder volume:

Cylinder volume

Max. rate

2 mL

6.67 mL/min

5 mL

16.67 mL/min

10 mL

33.33 mL/min

20 mL

66.67 mL/min

50 mL

166.00 mL/min

You can always enter a value between 0.01 and 166.00 mL/min

independent of the cylinder volume. When the function is carried out the

rate will be reduced automatically to the highest possible value.

3.8.7 Rotating direction of the valve disk (Dosing unit only)

Under Dosing unit/Valve disk you can specify the direction of

rotation of the valve disk. When choosing descending or ascending

respectively the ports will be moved at in descending or ascending

order. The shortest way results with the default setting automatic. If you

want to define a protected port, then choose not over i.e. the valve

disk will not be moved over this port. This is especially useful for

pipetting functions.