8 audit trail – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 46

Advertising
background image

3.7 System settings

34

PC Control / Touch Control

3.7.8 Audit

trail

In the Audit trail all user actions are recorded together with the user

name, date and time. Recording an audit trail is important for using PC

programs to meet the requirements of FDA Guideline 21 CFR Part 11

.

Under System/System settings/User administration in the

dialog Login options/Audit trail activate the checkboxes for

the data to be logged in the audit trail. The actions which should be

logged can be found in the online help. In order to meet the

requirements of FDA Guideline 21 CFR Part 11 for an audit trail

you must activate all the options.

You can also use the audit trail function specifically to record the data

which interest you.
To view the recorded data in PC Control you should open the menu

item Audit trail in the View menu.

Advertising