3 results, calculations and statistics – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 300

Advertising
background image

5.1 Troubleshooting

288

PC Control / Touch Control

5.1.3 Results, calculations and statistics

Problem

Possible cause

Remedy

The Statistics function has

not been activated in the

method under Sequence/

Method options.

Activate the statistics function

under Sequence/Method options

(see Section 3.16.4).

No Variable for mean has

been selected in the method

for the result calculation.

Select a Variable for mean in the

CALC command under Edit

result/Result options for the

result (see Section 4.4.1).

No statistics are

carried out for a

result.

The Statistics function is

switched off under Control.

Activate the Statistics function

under Control (see

Section 3.17.1).

Results have been

deleted in the

statistics, but the

variables (e. g. titer

or common variable)

have not been

reassigned.

If the statistics results are

subsequently altered then

the assignment is not

automatic.

Trigger the assignment of the

current statistics results with

[Recalculate] under Results (see

Section 3.18.8).

The numbering of the results

corresponds to the result

variables.

Under Results/Result

silo/Properties alter the settings

for the result silo display so that the

required result is shown (see

Section 3.22.1).

The result is not

shown in the result

silo.

The function Save result

in result silo has not

been switched on for the

result.

Activate this function for the

required result in the CALC

command under Result

options/More options (see

Section 4.4.1).

Advertising