4 – operation, Summary of operation steps – Drucker Diagnostics QBC STAR Dry Hematology Analyzer User Manual

Page 29

Advertising
background image

6000-300-000 4–1

4 – Operation

WARNINGS

BlOOd ANd BOdy fluIdS mAy cONtAIN the hepAtItIS B vIRuS (hBv),

hepAtItIS c vIRuS (hcv), humAN ImmuNOdefIcIeNcy vIRuS (hIv), OR OtheR

dISeASe-cAuSING AGeNtS. hANdle All pAtIeNt SpecImeNS AS pOteNtIAl

BIOhAzARdS cApABle Of tRANSmIttING INfectION. WeAR AppROpRIAte

peRSONAl pROtectIve equIpmeNt, INcludING lABORAtORy GlOveS,

WheN cOllectING, hANdlING, ANd pROceSSING BlOOd ANd BOdy fluIdS.

IN AddItION tO WeARING GlOveS, the uSe Of dISpOSABle lAB cOAtS OR

GOWNS ANd pROtectIve GlASSeS OR GOGGleS IS RecOmmeNded WheN

WORkING AROuNd the INStRumeNt.

AcRIdINe ORANGe ReAGeNt mAy Be tOXIc; dO NOt INGeSt. AvOId cONtAct

WIth SkIN, eyeS, ANd clOthING.

If A tuBe BReAkS IN the uNIt, cARefully RemOve the tuBe WIth A he-

mOStAt OR OtheR devIce, uSING puNctuRe ReSIStANt utIlIty GlOveS.

cONtAct techNIcAl SuppORt fOR AddItIONAl INfORmAtION.

Summary of Operation Steps

Routine operation of the QBC StaR Hematology System consists of the following steps:

• Turn the instrument on

• Print out / save electronic controls

• Fill tube with blood

• Mix the tube of blood

• Place the cap onto the tube

• Place the tube into the instrument

• Close door and ensure it is latched

• Push the “STAR” button to start the test cycle

• Obtain the results from the printer

• Dispose of the QBC StaR tube in a biohazard sharps container

Each of these steps is described below.

6000-300-000 4-1