Drucker Diagnostics QBC STAR Dry Hematology Analyzer User Manual

Page 60

Advertising
background image

QBC Star System Operator’s Manual

6000-300-000 7–10

417

Actuator position error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Place the instrument into Shipping Prep (Ref. page 2-9)
4. Power down the instrument (Ref. page 4-10)
5. Confirm the rotor spins freely by hand
6. If the rotor spins freely

a. Power down the instrument (Ref. page 4-10)
b. Turn the instrument Power on (Ref. page 4-2)
c. If the error persists contact QBC Technical Support

7. If the rotor does not spin freely, contact QBC Technical Support

418

Actuator current error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Place the instrument into Shipping Prep (Ref. page 2-9)
4. Power down the instrument (Ref. page 4-10)
5. Confirm the rotor spins freely by hand
6. If the rotor spins freely

a. Power down the instrument (Ref. page 4-10)
b. Turn the instrument Power on (Ref. page 4-2)
c. If the error persists contact QBC Technical Support

7. If the rotor does not spin freely, contact QBC Technical Support

419

Exhaust temperature sensor error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Tip the instrument up and check for obstructions in the

exhaust port

5. If there are obstructions, clear the obstructions and confirm

the proper setup (Ref. page 2-1)

6. Turn the instrument Power on (Ref. page 4-2)
7. if the error persists contact QBC Technical Support

420

Keypad error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power (Ref. page 4-2)
5. if the error persists contact QBC Technical Support

421

Run button error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power (Ref. page 4-2)
5. if the error persists contact QBC Technical Support

422

Lid latch release bar error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. If the lid is closed, open it manually (Ref. page 4-10)

a. If the lid opens, turn the instrument Power on (Ref. page 4-2)
b. If the error persists contact QBC Technical Support
c. If the lid does not open, contact QBC Technical Support

5. If lid is open, turn the instrument Power on (Ref. page 4-2)
6. If the error persists contact QBC Technical Support