4 – Operation

Start the test

To start the test, press the “STAR” button. The system automatically locks the door electrome-
chanically.

If there is no tube in the rotor or if the cap is not present, the message, “Sample Not Present
(open door)” appears. To proceed, place the tube in the instrument and press the “STAR”
button to resume testing.

The system can detect if a sample has already been tested. If this occurs, the message
“Press “STAR” button to re-scan sample” appears on the LCD. See below for instructions
on rerunning tubes.

The instrument mixes the tube contents for as long as 15 seconds using a low speed spin.
The float descends from the top of the tube toward the closure end of the tube. This mixes
the blood and reagents in the tube. A countdown displays the time remaining in the cycle.

After mixing is complete, the centrifuge accelerates to high speed to separate and pack
the cell populations into distinct cell bands. A countdown displays the time remaining in
the cycle.

Finally, a series of readings is taken. The messages “Reading Cycle In Progress,” is dis-
played while tests are being performed.

If an error message appears on the LCD display, please refer to section 7 (Troubleshooting)
for error resolution.

Rerunning tubes

the qBc StAR will process a tube that has been previously processed. to rerun a tube:

Insert a STAR Tube that has already been run into the STAR.

Close the Lid and press the run button.

The STAR will check the tube, and display the message “Press “STAR” button to re-scan sample”

Press the “STAR” button and the results are displayed on the LCD screen and internal printer.