6000-300-000 5–1

5 – Performance and Limitations

Operating Ranges

Hematology parameters measured with the QBC Star system are valid over the following range of values:

Hematocrit

15

65%

Hemoglobin

5.0

20.0

g/dL

Mean Corpuscular Hemoglobin Concentration

25 - 37.3 g/d

Platelet Count

20 – 999 x 10

9

/L

White Blood Cell Count

1.6 – 99.9 x 10

9

/L

Granulocyte Count

0.8 – 70.0 x 10

9

/L

Granulocyte

%

1

-

99%

Lymph/Mono Count

0.8 – 99.9 x 10

9

/L

Lymph/Mono

% 1-99%

results that fall outside these ranges may be confirmed by alternate methods.

Precision

Data on typical within-run precision tests on QBC Star tubes tested in the QBC Star system are shown
in the two tables below. the precision data represents the analysis of eleven whole blood specimens,
each assayed in replicates of 10.

Whole Blood total Imprecision

Parameter

Range

Max S. D.

GraN (%)

38 – 79

3.2

LYMPH/MONO (%)

21 – 63

3.2

Parameter

Mean Value

Mean %CV

HCt (%)

41.7

2.0 %

HB (g/dL)

14.0

1.9 %

PLt (x 10

9

/L)

235

6.0 %

WBC (x 10

9

/L)

6.0

6.4 %