Controls, Analytical quality control, Tri-level quality control label – Drucker Diagnostics QBC STAR Dry Hematology Analyzer User Manual

Page 45

Advertising
background image

6000-300-000 5–5

5 – Performance and Limitations

Controls

Analytical Quality Control

the QBC Star Centrifugal Hematology System has multiple built-in analytical quality control (QC)
systems that maintain the overall system integrity and the quality of the test results it produces.
the QBC Star System has five analytical quality control elements:

1. Factory calibration. System calibration is set during manufacture and cannot be altered by

the user.

2. Instrument Power On Self-test. this test assures that each time the instrument is turned

on, the computer, memory, optics, and motors are fully functional. Should you choose to
leave the system on continuously; the test will automatically be repeated every 8 hours if
the door is closed. a tri-level quality control label (QC label), designed to simulate 3 hema-
tology specimens (simulating low cell counts, normal cell counts, and elevated cell counts)
tests the system’s optics against values established at the time of manufacture. at the end
of the power on self-test, the instrument prints the values obtained from reading the QC
label. the values may be plotted to evaluate for shifts or trends in the data. the instrument
will flag any results that are outside the set limits, print an error code, and automatically
shut down operation of the instrument until the problem is corrected and a valid power on
self-test is performed.

3. Electronic QC (during each sample run). the built-in electronic checks during each sample

run confirm the proper centrifuge speed, centrifugation profile, system communications,
and internal temperature.

4. Sample Preparation QC (during each sample run). the built in checks confirm that the QBC

Star tube has not been previously processed. tests confirm that the tube assembly is
the proper length, the float is present and the correct length, and the tube is filled with the
correct amount of blood.

5. reagent QC (during each sample run). these built-in checks evaluate sample and reagent

integrity using the data from the optical scan. this includes tests for fluorescent signal
intensity, proper number, size and location of the cell layers and interface sharpness.

When these analytical quality control checks are successfully completed, the status of the instru-
ment’s analytical QC is printed on the patient record as “Star analytical QC: Passed.”results are
reported only if all of the analytical quality control requirements have been satisfied.

Tri-Level Quality Control Label

a tri-level quality control label, designed to simulate 3 hematology specimens, is automatically
read every 8 hours if the door is closed as part of the instrument power-on self-test. the instru-
ment prints the hematology values obtained from reading the QC label at the end of the power-on
self-test. these values are available to plot and evaluate shifts and trends.

the instrument compares the QC values measured (during power-on self-test to factory es-
tablished limits. results of any value outside of the established limits will result in an instru-
ment shutdown until the problem is corrected and a valid power-on self-test is performed.

6000-300-000 5-5

Advertising