Mix the tube – Drucker Diagnostics QBC STAR Dry Hematology Analyzer User Manual

Page 32

Advertising
background image

QBC Star System Operator’s Manual

6000-300-000 4–4

figure 14 – filling a tube from capillary Blood

For Capillary Blood Samples

Touch the collection tip of the STAR Tube to the sample while filling. The tube should be kept hori-
zontal or slightly inclined. Figure 14 shows a capillary blood sample.

For Venous and Capillary Blood Samples

Fill the QBC StaR Tube between the black fill lines. Blood must always be filled to at least the
first black line. This results in a blood volume of between 65 and 75 microliters. The instrument
automatically detects if the fill volume is correct.

mix the tube

Rock the QBC StaR Tube back and forth at least four times to mix the blood with the acridine
orange coating. Do not allow the blood to touch the white plug at the end of the tube.

Note: allowing the blood to flow from the collection end toward the plug end of the tube and back
is equal to one rock.

figure 15 – mixing the tube

6000-300-000 4-4