Printing of results – Drucker Diagnostics QBC STAR Dry Hematology Analyzer User Manual
Page 36
QBC Star System Operator’s Manual
6000-300-000 4–8
printing of Results
When the test is complete, the results are automatically displayed and printed.
NOte
dO NOt OpeN the INStRumeNt dOOR uNtIl the ReSultS ARe
RevIeWed ANd pRINted. test results are cleared from the display
when the instrument door is opened.
You must manually keep track of the link between the test results and the patient (unless
a keyboard is being used for patient identification). A space is provided on the printout for
writing the patient name or ID number. In addition, the date and time and the analytical QC
status (Pass, Failed) are printed.
the system only prints the actual measured or calculated results, with no interpretations
or normal values. the following hematology results are reported:
Reported Results
Parameter (abbreviation)
STD Units
SI Units
Hematocrit (Hct)
Percentage
Percentage
Hemoglobin (Hgb)
g/dL
mMol/L
Mean Corpuscular Hemoglobin Concentration (MCHC)
g/dL
mMol/L
White Blood Cells (WBC)
10
9
/L
10
9
/L
Granulocytes
10
9
/L, Percentage
10
9
/L, Percentage
Lymphocytes/Monocytes
10
9
/L, Percentage
10
9
/L, Percentage
Platelets (Plt)
10
9
/L
10
9
/L
Any results that are out of the reporting range (specified in Section 2) are shown on the display
with dashes instead of actual numerical values. On the printout, out of range readings are
indicated as greater than or less than the maximum or minimum value. Also the message
“Out of Operating Range” prints. Calculated values based on out of range readings print as
“No Report.” If the Analytical quality control verifications fail, an error message is displayed.
6000-300-000 4-8