6000-300-000 7–17

7 – Troubleshooting

10

Could not find closure

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid and visually check the sample

a. If the closure is not present, draw a new sample and retest
b. If the closure is present, rotate the sample 180 degrees

and rescan the sample

4. If the problem persists, create a new support file and contact

Technical Support

11

Could not find float bottom

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid and visually check the sample

a. If no visible buffy layer is observed, draw a new sample

and retest

b. If a visible buffy layer is observed, rotate the sample 180

degrees and rescan the sample

4. If problem persists, create a new support file and contact

Technical Support

12

Could not find RBC / Grans
interface

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid, rotate the sample 180 degrees and reinsert it

into rotor

4. Close lid and rescan the sample
5. If the problem persists, draw a new sample and retest
6. If the problem persists after retest, create a new support file

and contact Technical Support

13

Could not find Grans / LM interface

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid, rotate the sample 180 degrees and reinsert it

into rotor

4. Close lid and rescan the sample
5. If the problem persists, draw a new sample and retest
6. If the problem persists after retest, create a new support file

and contact Technical Support

14

Could not find LM / PLT interface

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid, rotate the sample 180 degrees and reinsert it

into rotor

4. Close the lid and rescan the sample
5. If the problem persists, draw a new sample and retest
6. If the problem persists after retest, create a new support file

and contact Technical Support

15

Could not find LM / Control interface

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid, rotate the sample 180 degrees and reinsert it

into rotor

4. Close the lid and rescan the sample
5. If the problem persists, draw a new sample and retest
6. If the problem persists after retest, create a new support file

and contact Technical Support