Replacing printer paper – Drucker Diagnostics QBC STAR Dry Hematology Analyzer User Manual

Page 49

Advertising
background image

QBC Star System Operator’s Manual

6000-300-000 6–3

1 Put on puncture-resistant gloves. Use a hemostat or other device to pick up any glass or plastic

fragments. Dispose of in a biohazard sharps container.

2 Clean any contaminated surfaces with 10% solution of household bleach (1 part bleach to 9

parts water). allow to stand 10 minutes, then rinse thoroughly with water and dry. Household
bleach is effective against bacteria, spores, and viruses. However, it is an oxidizing agent, and
is corrosive to metal alloys. Bleach must be thoroughly wiped off the instrument with a damp
cloth and dried. It should never be used if there is surface damage to any metal parts.

For any other cleaning methods, you must contact QBC Diagnostics to verify that the proposed method
does not damage the QBC Star instrument.

CAUTION

do not immerse the QbC STAR instrument in water or other liquid.

Replacing printer paper

When the printer paper is nearly exhausted, a colored edge appears on the roll of paper. You should
replace paper at this time to avoid running out while printing patient test results. See Figure 23.

To replace printer paper:

1 remove the printer panel and remove the old roll of paper.

2 Peel off the beginning of the new roll of paper.

3 Place the paper in the instrument, with the paper feeding up from the bottom.

4 Insert the paper between the two black rubber rollers on the printer.

5 Flip the paper release lever forward.

6 roll the manual paper advance wheel rearward. advance enough paper to clear the printer cover.

7 Flip the paper release lever backward.

8 replace the printer panel, feeding the new strip of paper through the slot.

figure 23 – Replacing printer paper

6000-300-000 6-3

Figure 23.