Removing installed probes – NARDA 8718B User Manual

Page 109

Advertising
background image

Operating the Meter

108

7. Enter any special calibration frequencies on the extra

lines provided and enter the appropriate calibration

factors. Delete any standard calibration frequencies

that were not used.

8. Click on the Install Probe button. A dialog box will

appear asking you if you are sure you want to install

the probe.

9. Click on OK. The program will automatically send the

calibration information to the meter via the modem and

cable. The program will then verify that the probe has

been successfully installed in the meter by reading the

meter’s memory.

10. The model number and the serial number of the probe

just installed will now appear in the table of Installed

Probes.

11. To leave the Probe Installation window and return to

the main menu click the Close button.

Removing Installed Probes

To remove a specific probe from the meter’s memory...

1. First view an installed probe as described in Viewing

Installed Probes.

2. Click on the Remove Probe button. A dialog box will

appear asking you to confirm that you want to remove

a probe from the meter’s memory.

3. Click OK. The probe will automatically be removed

from the meter’s memory.

4. The model number and the serial number of the probe

just removed will disappear from the table of Installed

Probes.

5. To leave the Probe Installation window and return to

the Main Menu Window click the Close button.