The readout on the 8718B is shown in terms of Percent of

Standard. The full scale of Narda’s shaped probes is typically

either 300% or 600% of a standard. This permits measurements

of exposure environments where the weighted field strengths

range from as high as three to six times the maximum

permissible exposure (MPE) level to environments where the

levels are less than one percent of the MPE.

Connecting a Probe

Connect a cable with a quick-release, eight-pin connector to the

connector at the top of your Model 8718B. This cable will be

either (1) a short adapter cable for attachment to older style

8700 series probes with integral cables or (2) a 4 ft. (1.2m) long

cable with a quick-release connector on both ends for use with

the 8700 "D" series probes. The D series probes have a

quick-release, eight-pin connector that allows the probe to

attach directly to your 8718B or be used with a cable

A direct connection can be useful for certain low frequency

measurements. See Low Frequency Measurements in the

Applications Booklet.

8718B

8718B

8718B

8700D Series Probe

8700D Series Probe

8700/8700B Series Probe

Direct Connection

4' (1.2m) cable
Model 8744-04 (supplied)

1' (0.3m) adapter cable
P/N 2178770 (optional accessory)

It is important to un-

derstand that each

shaped frequency re-

sponse probe model

is designed

to conform to a par-

ticular standard.