Chapter 4

Making a Measurement

Measurement Uncertainty

The uncertainty of any measurement includes several factors

related to the probe and meter.

Frequency response or frequency sensitivity is typically

±1 dB to ±2 dB (from about ±25% to about ±55%).

The correct use of correction factors can greatly reduce

the amount of uncertainty associated with frequency

response.

Checking or Changing the Probe Model

To check the probe model that the meter is currently set to...

1. Press F3 (MENU). The First Menu Screen will appear.
2. Press 7 (PROBE). A Screen will appear that shows the

model number and basic characteristics of the probe.

The probe serial number is shown for probes selected

from memory.

3. Press F1 (OK) key to return to the First Menu Screen.
4. Press Esc to return to the Measurement Display Screen.

To change to another probe without turning the meter off...

1. Press F3 (MENU). The First Menu Screen will appear.
2. Press 7 (PROBE). A screen will appear that shows the

model number and basic characteristics of the probe.

3. Press F4 (RESTART) to return to the Opening Menu

Screen. Press F1 (MEASURE) and begin the probe

selection process.