Chapter 14 Probe Zero

Zeroing the Probe While In Use

The meter and probe are initially zeroed before the meter is

allowed to go into the measurement mode. If you determine that

significant drift may have occurred, it is simple to fine tune the

zero. See Checking for Zero Drift on page 82.

To zero the probe while in use...

1. Shield the probe using one of the methods described

in Shielding the Probe.

2. Press the Auto Zero key. The meter will correct for any

probe drift in about eight seconds. The meter will then

automatically return to the measurements screen.