Chapter 4
Making a Measurement
To select from the PROBES IN MEMORY...
1. If you haven’t already, press F1 (MEASURE).
2. The model number and the serial number of the last
probe used will be displayed on the second line of the
display. If you want to use the same probe, press
Enter.
3. If you want to use a different probe, look for the model
number in the positions above the F1, F2, or F3
legends. If more than three probes are in memory they
can be found by pressing F4 (MORE).
4. Press the appropriate function key and the probe
model and serial number will be displayed on the
second line of the screen. If this is correct, press Enter.
5. A new screen will appear specifying probe model
number, serial number, frequency range, field type,
maximum measurement range, and type of response
(flat or shaped). If this information is correct, press F1.
If this information does not match your requirements,
you need to select another probe. Press F2 and you
will go back to the Select Probe screen to make
another selection.
6. Proceed with the probe zeroing process. See Zeroing
the Probe on page 17.
S E L E C T
P R O B E
M O D E L
A 8 7 2 2
S N
D
X X X X X
O R
S E L E C T
F R O M
L I S T
8 7 X X
8 7 X X
8 7 X X
M O R E
F1
F2
F3
F4
A 8 7 4 2 D
S / N
0 0 0 0 1
3 0 0
H
- 3 G H z
z
k
E - F L D
6 0 0
%
S T D
S H A P E D
Y E S
N O
M E N U
F1
F2
F3
F4