Alarm settings, Alarm description, Controlling the alarm functions – NARDA 8718B User Manual
Page 68
67
Chapter 10 Alarm Settings
10
Alarm Settings
Alarm Description
The Model 8718B Meter is equipped with an audible alarm that
emits sound through a tiny hole on the bottom of the meter. The
alarm sounds automatically whenever the field strength
exceeds the measurement range of the probe or over ranges.
The meter can also be set to sound its alarm at a specific level
and/or to sound a variable tone indication of field strength.
Controlling the Alarm Functions
To make changes to the audio alarm settings...
1. Access the menu system by pressing the function key
that corresponds to the legend below MENU on the
bottom line of the display.
2. Press 8 (NEXT) to go to the Second Menu Screen.
3. Press 5 (ALARM) to access the alarm menu. A new
menu will appear that allows you to:
•
Turn the audio alarm on or off.
•
Set the alarm threshold in terms of percent of full scale
of the probe in use (1% to 99%).
•
Turn the variable tone alarm on or off.