System checks – NARDA 8718B User Manual

Page 99

Advertising
background image

98

Operating the Meter

8. Press F4 (CABLE) to change to the cable input. The

meter will return to the First Menu Screen. Press Esc to

return to the measurements screen.

System Checks

To insure that the entire system – meter, fiber optic

transmitter, and probe – are functioning properly...

1. Set the 8718B to use the fiber optic receiver input as

previously described.

2. Connect the probe to the fiber optic transmitter.
3. Connect a fiber optic cable between the meter and the

fiber optic transmitter.

4. Check the battery charge level in the transmitter by

turning switch S1 (BATT TEST–ON) to BATT TEST.

Verify that the green LED is illuminated which signifies

that the battery is adequately charged for operation. If

the LED fails to illuminate, charge the transmitter

overnight. A full-battery charge takes approximately 16

hours.

5. Set S1 to the ON position and S2 to the NORM

position.

6. Zero the 8718B using the standard procedure.
7. Move S2 ( CAL1NORMCAL2) to the CAL1 position.

The meter should read 45-55% of the full scale rating of

the probe. The actual error from the combination of the

8718B and the fiber optic transmitter is less than (±3%).

8. Move S2 to the CAL2 position. The meter should read

approximately 25% higher than the CAL1 position. The

exact value is not critical.

9. Set S2 to the NORM position to begin making

measurements providing that all the system checks

were successfully completed.