Using rs-232 – Xantrex Technology XDC 20-600 User Manual

Page 132

Advertising
background image

Remote Operation

4–18

TM-XDOP-01XN

SCPI

Set the band rate:

SYST:COMM:SER:BAUD {1200 | 2400 | 4800 | 9600 | 19200 |

38400}

Select flow control:

SYST:COMM:SER:PACE {HARD | XON | NONE}

Where

HARD means hardware flow control

XON means XON/XOFF characters are used

NONE means no flow control

Select RS-232 to be the remote control source:

SYST:REM:SOUR RS232

Put the unit into remote operation:

SYST:REM:STAT REM

Using RS-232

Use any terminal emulation program to send commands to the power supply.

To change between remote and local modes, press the LCL/RMT button or the
command:

SYST:REM:STAT {REM | LOC | RWL}

If you are in local mode, you may still communicate receive responses to queries,
but you may not change any settings. Attempting to do so will cause Error -221,
"Settings conflict".