Setting up current sharing – Xantrex Technology XDC 20-600 User Manual

Page 189

Advertising
background image

Setting up Current Sharing

TM-XDOP-01XN

5–3

Setting up Current Sharing

To set up multiple supplies for current share operation:

1. Configure each supply with a unique multichannel address. See

“Multichannel Operation” on page 4–11.

2. Configure the supplies for current sharing. Configure one supply to operate as

the master, the others as slaves.

Current sharing may be configured either through the front panel menus or by
SCPI commands sent via a remote interface.

Front Panel

In addition to selecting the current share mode, the master may be set up to
display the summed current from all units.

On the front panel, a

MASTER

or

SLAVE

annunciator will light up to indicate the

current share mode.

To configure current share with the front panel:

.

SCPI

Select whether the unit will operate as a master or slave unit:

SOURCE:COMBine:CSHare:MODE [MASTer | SLAVe | OFF]

To query the total output current of all current sharing units, use the SCPI
command:

MEAS:CURR? SUM

The current share subregister (CSHare) will show whether the master or slave is
operating.

Step # Do This

You Will See

1

ERROR MSGS

2

or

Press 8 times

CURRENT SHARE

3

No share

4

or

to select the current

share option (No share, Master or Slave).

Press

to

save.

If No share or Slave are selected, configuration is complete and the menu exits.

Master

4a

If Master has been selected, you will be prompted to select whether the master
should display the total current output of all current share units.

Use

or

to select Y or N.

Press

Display Sum? Y

MENU

MENU



ENTER





ENTER





ENTER