Configure remote control source – Xantrex Technology XDC 20-600 User Manual

Page 82

Advertising
background image

Operation

3–20

TM-XDOP-01XN

Configure Remote Control Source

The REMOTE CONFIG menu option lets you set up the attributes of the remote
control sources.

The following table shows how to access and work with the Remote
Configuration option.

After you have selected the remote control source you want to configure,
configure that setting using appropriate values from the following table. Select
values using the arrow keys or the numeric keypad. To save a value you have
selected, press ENTER.

See Chapter 4, “Remote Operation” for detailed instructions on setting up remote
interfaces.

Step # Do This

You Will See

1

ERROR MSGS

2

5 times

a

a.

You can also press MENU 5 times.

REMOTE CONFIG

3

RS-232 Cfg

4

or

Press repeatedly until the desired
remote control source appears.

RS-232 Cfg
Analog Cfg
GPIB Cfg
Multichnl Cfg

5

See the tables that follow to see
how to proceed.

MENU



ENTER





ENTER

Remote Control
Source

Prompt

Configuration Settings

RS-232

Baud ######

Flow Ctl ####

1200, 2400, 4800, 9600, 19200, 38400

a

Hdwr, XON, None

b

Analog

Input #### V

Select the operating range of the API:
0–5
0–10

GPIB

GPIB Addr ##
PON SRQ? Y

1–30
Y, N

Multichannel

Slave Addr ##
Connect?

2-50
Y, N

a.

This range depends on the user’s network configuration.

b.

Hdwr = hardware handshake; XON = software flow control XON/XOFF characters used;
None = no flow control

TM-XDOP-01XN.book Page 20 Monday, July 17, 2006 11:19 AM