Remote Operation

4–34

TM-XDOP-01XN

Configure Auxiliary Status Lines

To set up Auxiliary line A:

OUTP:AUXA:SOUR <aux-line-mnemonic>

The choices for <aux-line-mnemonic> include:

NONE

ON: Output On

OFF: Output Off

OVOL: Over-Voltage Condition

UVOL: Under-Voltage Condition

OCUR: Over-Current Condition

UCUR: Under-Current Condition

OPOW: Over-Power Condition

UPOW: Under-Power Condition

ACOF: AC Power had been turned off or failed

OTEM: Over-Temperature Condition

HTEM: High-Temperature Condition

SPR: Sense Protection Tripped

FOLD: Fold Protection Tripped

CC: Constant Current Mode

CV: Constant Voltage Mode

CP: Constant Power Mode

UNR: Output Unregulated

To check the setting:

OUTP:AUXA:SOUR?

To set up Auxiliary line B:

OUTP:AUXB:SOUR <aux-line-mnemonic>

To set the polarity of the auxiliary status lines:

OUTP:AUXA:POL {HIGH|LOW}

OUTP:AUXB:POL {HIGH|LOW}

Where:

HIGH

means that the logic of the output is active high. (That is, if the condition is

true, the line is pulled high.)

LOW

means the logic of the output is active low.