Query error list, User request event, Operation complete event – Xantrex Technology XDC 20-600 User Manual

Page 264

Advertising
background image

Error Messages

D–6

TM-XDOP-01XN

Query Error List

An error number in the range [-499, -400] indicates that the output queue control
of the instrument has detected a problem with the message exchange protocol
described in IEEE 488.2, chapter 6. The occurrence of any error in this class
causes the query error bit (bit 2) in the Event Status Register to be set.

User Request Event

An error/event in the range [-699, -600] is used when the instrument wishes to
report a 488.2 user request event. This event also sets the user request bit (bit 6) of
the Standard Event Status Register.

Operation Complete Event

An error/event in the range [-899, -800] is used when the instrument wishes to
report a 488.2 operation complete event. This event occurs when an instrument’s
synchronization protocol, having been enabled by an *OPC command, completes
all selected pending operations. This event also sets the operation complete bit (bit
0) of the Standard Event Status Register.

Table D-4 Query Error List

Error code Error Message Description

-400

Query Error
This is the generic query error for the power supply, used only when the
other types of errors do not apply.

-410

Query INTERRUPTED
Generated when a new command was received before it could finish the
query.

Table D-5 User Request Event

Error code Error Message Description

-600

User request

Table D-6 Operation Complete Event

Error code Error Message Description

-800

Operation complete

TM-XDOP-01XN.book Page 6 Monday, July 17, 2006 11:19 AM