Xantrex Technology XDC 20-600 User Manual

Page 153

Advertising
background image

SCPI Commands for Digital Interfaces

TM-XDOP-01XN

4–39

FORever and INFinity cause the sequence to repeat forever. A query will
return 9.9E37, representing INFinity.

Selecting a Trigger Source

If any triggers are programmed into the sequence, select a trigger source:

PROG:TRIG:SOUR {BUS|MAN|EXT|IMM}

BUS

- trigger signal is IEEE 488.1 GET or *TRG

MANual

- trigger input is from the front panel TRIGGER key

EXTernal

- The external trigger line is selected as the source.

IMMediate

- The source is the SCPI command “INIT:IMM”

Editing Step Parameters

Commands are provided to edit only one of voltage, current, power, OVP level,
step duration, or end action.

Program step voltage:

PROG:STEP<step_number>:VOLT <voltage>

Program step current:

PROG:STEP<step_number>:CURR <current>

Program step power:

PROG:STEP<step_number>:POW <power>

Program step OVP level:

PROG:STEP<step_number>:OVP <OVPlevel>

Program step time (duration):

PROG:STEP<step_number>:DWEL {<time>|TRIG}

To delete a step:

PROGram:STEP<step_number>:DELete

An error will result if the last step with the end action is deleted, without a new
end action programmed to take its place.

Deleting Sequences

If you no longer need a sequence, select it with the PROG:NAME command and
then delete it with the following:

PROGram:DELete

You can also use the following to delete all sequences:

PROGram:DELete:ALL

TM-XDOP-01XN.book Page 39 Monday, July 17, 2006 11:19 AM