Remote Interface Calibration Procedure

TM-XDOP-01XN

B–13

Analog Programming Interface Voltage Readback Calibration

1. Attach a DVM across the voltage readback lines, Pins B5 and B1(GND).

2. Minimum calibration level

Set the voltage readback lines to approximately 10% of full scale by sending
the command:

CAL:ANAL:5V:READ:VOLT:LEV MIN

3. Enter voltage data

Enter the voltage across the voltage readback lines, read from the external
DVM.

CAL:ANAL:5V:READ:VOLT:DATA <voltage>

4. Maximum calibration level

Set the voltage readback lines to approximately 90% of full scale by sending
the command:

CAL:ANAL:5V:READ:VOLT:LEV MAX

5. Enter voltage data

Enter the voltage across the voltage readback lines, read from the external
DVM.

CAL:ANAL:5V:READ:VOLT:DATA <voltage>

6. Power supply calculates and stores calibration constants.

Analog Programming Interface Current Programming Calibration

1. Connect the power source across the current programming lines, Pins B4 and

B1(GND). Attach a DVM across the current programming lines as well.

2. Minimum calibration level

Set the input to the programming lines to approximately 0.5V (10% of full
scale).
Set the power supply to receive 5V analog current programming calibration
data with the command:

CAL:ANAL:5V:PROG:CURR:LEV MIN

3. Enter voltage data

Enter the voltage at the current programming lines, read from the external
DVM.

CAL:ANAL:5V:PROG:CURR:DATA <voltage>

4. Maximum calibration level

Set the input to the programming lines to approximately 4.5V (90% of full
scale).
Set the power supply to receive 5V analog current programming calibration
data by sending the command:

CAL:ANAL:5V:PROG:CURR:LEV MAX