Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1041
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1025
EP criterion
Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break points will be found. As this is a relative value
related to the total measured value modification, even small changes in
the measured value can be evaluated as a break point for a small meas-
ured value range.
Range
0 ... 1.0
Default value
0.3
Slope
Minimum difference between the slope before and after the break point.
The smaller the difference, the more break points will be found.
Range
0.0 ... 10.0
Default value
0.9
Smoothing factor
The higher the smoothing factor, the fewer endpoints will be found.
Range
2 ... 20
Default value
5
Window
Selection
Measured value | Time | off
Default value
off
off
A range (window) can be defined on the measured value axis or on the
time axis. The break point evaluation will only be carried out in the
defined window. Only the first break point in the defined window will
be recognized.
Lower limit
Value for the lower limit of the window.
Time for the lower limit of the window.
Window = Measured value
Range
0.0000 ... 2000.0 mS/cm
Default value
0.0000 mS/cm
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.