5 set u - titration parameters, Titration parameters – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 747
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
731
EP2 at pH
Measured value for the second endpoint.
Titrando, 855
Range
-20.0 ... 20.0
Selection
off
Default value
off
Titrino
Range
-20.00 ... 20.00
Selection
off
Default value
off
For additional parameters for Endpoint 2, see Endpoint 1.
5.6.3.4.4.5
SET U - Titration parameters
Tab: Method
▶ SET U ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.
Titration direction
Selection of the titration direction. If two endpoints are set, then the titra-
tion direction is already established and the following selection of titra-
tion direction is ignored.
Selection
+ | - | auto
Default value
auto
+
positive measured value alteration
-
negative measured value alteration
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.
Extraction time
The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). For example, it makes sense to
enter an extraction time for the titration of samples that are not easily
soluble.