Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1649

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Instruments

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1633

of the actuating variable which a PD controller causes immediately as a
result of the D component.

for control task = pump control
Range

0 ... 9999.0

Default value

0.0

for control task = free controller
Range

0.0 ... 9999.0 s

Default value

0.0 s

Attenuation time

for control task = pump control
Range

0.0 ... 9999.0

Default value

0.0

for control task = free controller
Range

0.0 ... 9999.0 s

Default value

0.0 s

D in feedback path

on | off (Default value: off)

Upper limit

Upper limit value at which the integration is stopped and the output is
limited.

Range

0% mapping of the output ... 100% mapping
of the output 'Mapping unit of the output'

Default value

100% mapping of the output 'Mapping unit
of the output'

Lower limit

Lower limit value at which the integration is stopped and the output is
limited.

Range

0% mapping of the output ... 100% mapping
of the output 'Mapping unit of the output'

Default value

0% mapping of the output 'Mapping unit of
the output'

ARW only with I component

on | off (Default value: off)

Advertising