6 properties - electrode test – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1448
6.7 Subwindow Sensors
■■■■■■■■■■■■■■■■■■■■■■
1432
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
[Limits]
Open the dialog window Limits for sensor, in which the warning and
intervention limits for the calibration data can be defined. These limits only
apply to the graph, no monitoring will take place during calibration. For
monitoring during calibration, the limits set on the Limits tab apply.
[Delete history]
Delete the sensor history.
6.7.3.6
Properties - Electrode test
Dialog window: Configuration
▶ Sensors ▶ Edit ▶ Properties ▶ Electrode test
Date ELT
Date of the last electrode test. The date can be selected by pressing
in
the dialog window Select date. This date is used for the electrode test
monitoring (Date ELT + Interval ELT = Next ELT).
Result
Indicates the quality of the tested electrode.
Selection
Excellent | Good | Usable | Not tested
Default value
Not tested
Monitoring the ELT validity
on | off (Default value: off)
If this check box is activated, then the time interval for the electrode test is
monitored.
Can be edited only if a date has been entered in the field Date ELT.
Interval ELT
Time interval to next electrode test. If a value is entered here then the
date in the field Next ELT will be adjusted automatically.
Can be edited only for Monitor ELT validity = on, otherwise inactive.
Range
1 ... 999 Days
Default value
999 Days
Next ELT
Date on which the next electrode test is to be carried out. The date can be
selected by pressing
in the dialog window Select date. Once the date
has been entered, the ELT field will be adjusted automatically.
Can be edited only for Monitor ELT validity = on, otherwise inactive.
Default value
Date ELT+ 999 days