5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1229

Device

Device name

Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display of the device type. If a device is selected under Device name,
then this Device type field can no longer be edited, and the device type
belonging to the selected device is displayed instead.

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

774 Oven Sample Processor | 089 Photometer
| 874 Oven Sample Processor

Default value

874 Oven Sample Processor

Heater (Sample changer)

Target temperature

Temperature to which the oven has to be heated. This value is live edita-
ble only for the 874.

Range

50 ... 250 °C

Selection

Init.temp. | off

Default value

Init.temp.

Init.temp.
With Init.temp. the oven heating is switched on at the same time as
the device and the oven is heated up to this temperature. The initial
temperature must be entered by keypad on the 774. When connecting
to tiamo, the value is read out and displayed in the device configura-
tion. The value cannot be altered in tiamo. The initial temperature is
entered and altered in tiamo with the 874.
off
The heater is switched off during the determination.