5 kfc - titration parameters, 6 kfc - stop conditions, Titration parameters – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 854: Stop conditions
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
838
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
5.6.3.6.5
KFC - Titration parameters
Tab: Method
▶ KFC ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.
Extraction time
The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.
Range
0 ... 999999 s
Default value
0 s
Temperature
Manually entered titration temperature.
756, 831
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
851, 852
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
756, 831
Range
1 ... 999999 s
Default value
2 s
851, 852
Range
0.1 ... 999999.0 s
Default value
2.0 s
5.6.3.6.6
KFC - Stop conditions
Tab: Method
▶ KFC ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters