5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
883
Wait for [Continue]
Reagent dosing in the running STAT command will be interrupted and
a message will be displayed. As soon as the monitored measured value
is again within the limits (including hysteresis) reagent dosing can be
resumed by pressing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command will be interrupted. As
soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.
Note
If the Lower limit is infringed while monitoring the dosing rate, the
actions Wait for [Continue] and Wait for limit ok do not make
sense because the average dosing rate becomes smaller and smaller
during the waiting time.
Monitoring Temperature
Monitoring Temperature
on | off (Default value: off)
If this option is activated, then the temperature will be monitored and any
exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the temperature. If the temperature falls below this limit,
the event Lower temperature limit exceeded is triggered.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
-20.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
-170.0 °C
Lower hysteresis
Lower hysteresis of the temperature. If the temperature once again
exceeds the lower limit by this hysteresis value, then the event Tempera-
ture lower limit ok is triggered.