5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

486

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

9999.99 mL

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.

Titrando, 855
Range

0 ... 200

Default value

5

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.2.2.11

DET pH - Fixed endpoint evaluation

Dialog window: Method

DET pH Properties... Additional evaluations Fixed

endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value