11 det ph - fixed endpoint evaluation – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 502
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
486
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.
Titrando, 855
Range
0 ... 200
Default value
5
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.2.2.11
DET pH - Fixed endpoint evaluation
Dialog window: Method
▶ DET pH ▶ Properties... ▶ Additional evaluations ▶ Fixed
endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value