5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1131

from other measuring command) by which every measured absorbance
value can then be corrected.

Range

-4000 ... 4000 mAU

Default value

0 mAU

Integration time

Integration time for adsorption measurement.

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.

Range

1 ... 10000

Default value

10

Smoothing

Number of neighboring pixels with which the light intensity is determined
for each measurement pixel.

Range

1 ... 100 Pixels

Default value

1 Pixels

Parameters for the 089 Photometer

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.

Range

0 ... 4000 mAU

Default value

0 mAU

Measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control
can be selected.

Selection

Measurement with drift control | Measure-
ment without drift control

Measurement with drift control
The measurement is carried out drift-controlled if this option is selec-
ted. The measurement is canceled as soon as the defined Signal drift
or the Stop measured value is reached or a Max. waiting time has
elapsed. The corresponding parameters are not visible if this option is
disabled.