5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1131
from other measuring command) by which every measured absorbance
value can then be corrected.
Range
-4000 ... 4000 mAU
Default value
0 mAU
Integration time
Integration time for adsorption measurement.
Range
0.01 ... 600000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are recorded and averaged.
Range
1 ... 10000
Default value
10
Smoothing
Number of neighboring pixels with which the light intensity is determined
for each measurement pixel.
Range
1 ... 100 Pixels
Default value
1 Pixels
Parameters for the 089 Photometer
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Range
0 ... 4000 mAU
Default value
0 mAU
Measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control can be selected.
Selection
Measurement with drift control | Measure-
ment without drift control
Measurement with drift control
The measurement is carried out drift-controlled if this option is selec-
ted. The measurement is canceled as soon as the defined Signal drift
or the Stop measured value is reached or a Max. waiting time has
elapsed. The corresponding parameters are not visible if this option is
disabled.