7 Instruments

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1625

Unit

Unit of the measured value to be used.

Can be edited only for controlled variable = calculated value, other-
wise grayed-out and not available for editing.

Input

50 characters

Default value

% or controlled variable unit, respectively

Default value

Selection or input of the standard value which is to be used for the input if
the result is invalid.

Range

0% mapping of the input ... 100% mapping
of the input

Selection

last value | setpoint value

Default value

last value

Setpoint value

Setpoint value to which the control is to be set.

Range

0% mapping of the input ... 100% mapping
of the input 'Mapping unit of the input or of
the calculated value'

Default value

0 'Mapping unit of the input or of the calcula-
ted value'

Hysteresis

Hysteresis for setpoint value.

Range

0 ... Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the analog input or of the calcu-
lated value'

Default value

0 'Mapping unit of the analog input or of the
calculated value'

Actuating variable

Selection of the actuating variable out of all of the digital outputs of the
Controller output type.

Selection

'Port description'

'Port description'
Only digital output ports of the Controller output type can be selec-
ted.